Repository landing page

We are not able to resolve this OAI Identifier to the repository landing page. If you are the repository manager for this record, please head to the Dashboard and adjust the settings.

The New Generation of the KIT 3D USCT

Abstract

The first clinical studies with our current prototype, 3D USCT II, enabled us to identify the necessary improvements for transition of our method to clinical practice. The main goals are to improve the contrast of reflection and transmission tomography, and to optimize the coverage of the imaged breast by a new geometry of the transducer distribution. Furthermore, for cost-effective industrial mass production, a self-calibration method allows us to relax the precision of the positioning of the transducers to 0.1 mm. The readout of the transducer arrays is now carried out by an ASIC, developed for a more cost-effective design. The coupling of the measuring device to the patient was optimized to cover the full size of the breast up to the pectoral muscles. Finally, the data acquisition and readout time were reduced to 1.5 minutes each by new micro-TCA electronics and larger FPGAs

Similar works

This paper was published in KITopen.

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.