Wide field epiretinal micro-electrode-design and feature test

Abstract

This study was aimed to design and fabricate a wide field implantable epi-retinal microelectrode array for the purpose of retinal repair, and to perform electrochemical test on the array. With parylene as flexible substrate material and Pt as electrode and route material, microelectrode array prototypes was designed and fabricated, and electric characteristics of the array was tested with the three-electrode test system. The feature analysis showed that morphological and electrical properties of the array well met the requirements of implantation and electrical stimulation of retina. The microelectrode array can be put in the in vivo electrophysiological experiments on animal and can perform reliably. © 2013 IEEE.published_or_final_versio