Integrated multipoint-laser endoscopic airway measurements by transoral approach

Abstract

Objectives: Optical and technical characteristics usually do not allow objective endoscopic distance measurements. So far no standardized method for endoscopic distance measurement is available. The aim of this study was to evaluate the feasibility and accuracy of transoral airway measurements with a multipoint-laser endoscope. Methods: The semirigid endoscope includes a multipoint laser measurement system that projects 49 laser points (wavelength 639 nm, power < 5mW) into the optical axis of the endoscopic view. Distances, areas, and depths can be measured in real-time. Transoral endoscopic airway measurements were performed on nine human cadavers, which were correlated with CT measurements. Results: The preliminary experiment showed an optimum distance between the endoscope tip and the object of 5 to 6 cm. There was a mean measurement error of 3.26% ± 2.53%. A Spearman correlation coefficient of 0.95

Similar works

Full text

thumbnail-image

Qucosa - Publikationsserver der Universität Leipzig

redirect
Last time updated on 03/09/2019

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.